Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

First implementation of patients synchrony #59

Open
wants to merge 1 commit into
base: develop
Choose a base branch
from

Conversation

jast1982
Copy link
Contributor

@jast1982 jast1982 commented Jun 3, 2020

created a virtual inhalation sensor out of pressure sensor input
A sensor value > threshold detects a patients breath and directly allows
to go from exhalation to inhalation. Timing for subsequent breaths is
adjusted put pulled back 20% per breath if patient is not triggering
another breath

  • max pressure is taken from parameters

created a virtual inhalation sensor out of pressure sensor input
A sensor value > threshold detects a patients breath and directly allows
to go from exhalation to inhalation. Timing for subsequent breaths is
adjusted put pulled back 20% per breath if patient is not triggering
another breath
- max pressure is taken from parameters
@jast1982 jast1982 requested a review from mrrosen June 3, 2020 15:35