prasimix commented Jun 15, 2017

If current auto-range is set to auto (SENS:CURR:RANG:AUTO ON), and current range is set to best/default (SENS:CURR:RANG DEF) when output is in unregulated mode (OUTP:MODE? is UR, that can happen on some Power boards when voltage is set to small values) and current is set above 0.5 A firmware will starts to switch repeatedly current range between low and high range.

@prasimix prasimix added this to the v1.0 milestone Jun 15, 2017
@prasimix prasimix changed the title Switching between low and high current range for low set voltage Switching current ranges if low output voltage is set Jun 15, 2017
mvladic added a commit that referenced this issue Jun 15, 2017
@prasimix prasimix changed the title Switching current ranges if low output voltage is set Continuous switching current ranges when output is in UR mode Jun 15, 2017
mvladic added a commit that referenced this issue Jun 16, 2017
mvladic added a commit that referenced this issue Jun 16, 2017
mvladic added a commit that referenced this issue Jun 16, 2017
@prasimix
Copy link
Member Author

Fixed