Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

SYSTem:COMMunicate:ENABle does not affect displayed icon #172

Closed
prasimix opened this issue Oct 1, 2017 · 2 comments
Closed

SYSTem:COMMunicate:ENABle does not affect displayed icon #172

prasimix opened this issue Oct 1, 2017 · 2 comments
Assignees
Milestone

Comments

@prasimix
Copy link
Member

prasimix commented Oct 1, 2017

SYSTem:COMMunicate:ENABle command does not affect displayed icon.

For example if SYSTem:COMMunicate:ENABle 0, ETH is executed, Ethernet communication will be disabled but its icon will not be changed. If that is related to required restart of the PSU, there is no reason why we shouldn't display icon in "error" state (red background).

Additionally it seems that SYSTem:COMMunicate:ENABle 0, SER doesn't work at all. No icon change or port status can be found under Serial (via USB) settings.

@prasimix prasimix added this to the v1.02 milestone Oct 1, 2017
mvladic added a commit that referenced this issue Oct 4, 2017
@prasimix
Copy link
Member Author

prasimix commented Oct 4, 2017

If Ethernet is disabled (no icon is shown on local display), sending of SYSTem:COMMunicate:ENABle 0, ETH will display red Ethernet icon, what is not correct.

mvladic added a commit that referenced this issue Oct 4, 2017
@prasimix
Copy link
Member Author

prasimix commented Oct 4, 2017

Fixed

@prasimix prasimix closed this as completed Oct 4, 2017
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment
Projects
None yet
Development

No branches or pull requests